TURBINE FLOW METER

[SERIES : TFM-011]

DESCRIPTION:

A turbine flow meter is constructed with rotor and blades that use the mechanical energy of the fluid to rotate the rotor in the flow stream. Blades on the rotor are angled to transform energy from the flow stream into rotational energy. The rotor shaft spins on bearings: when the fluid moves faster, the rotor spins proportionally faster. Shaft rotation can be sensed mechanically or by detecting the movement of the rotor blades. Rotor movement is often detected magnetically, where movement of the rotor generates a pulse. When the fluid moves faster, more pulses are generated. Turbine flow meter sensors detecting the pulse are typically located external to the flowing stream to avoid material of construction constraints that would result if wetted sensors were used. The RPM of the turbine wheel is directly proportional to the mean flow velocity within the tube diameter and corresponds to the volume flow over a wide range.

FEATURES:

  • TFM-011
  • Simple & cost-effective
  • Suitable for non-conductive clear liquids
  • Local indication through LCD
  • Durable & versatile
  • Easy maintenance
  • Die cast aluminium IP-68 housing
  • Remote display optionally available

Installation Specifications

MediaLiquids (Clear)
Viscosity100 cp max
Line Size15 NB to 150 NB
Display16X2 LCD-4 digit for flow rate & 8 digit for totalized flow (flow rate & totalized flow will be visible through toggle key)
Transmitter ElectronicsOption 1: Integral
Option 2: Remote Remote Electronics Cable: 25mtr
Type of Output1) 4-20mA DC
2) Pulse-30mA Peak-Peak
Calibration RangeAs per requirement (Factory Calibrated)
Accuracy±1% of F.S. (For 20-100% flow)
Linearity±1%
Repeatability±1%
Pressure DropApprox. 0.28 kg/cm² at max flow
Turn down ratio10:1 or 100:1
Process Temperature0-85°C
Max Process Pressure0-10 kg/cm²
Material of constructionBearings: Tungsten Carbide Sleeve/V Jewel
Rotor: SS410/17.PH
Shaft: Tungsten Carbide
Body/Support/Flange: SS
Power Supply24 V DC
Power Consumption<0.5 VA
Response Time<100mSec
Temperature coefficient±0.01% per degree C
Transmitter EnclosureDie Cast Aluminum IP 68
Process Connections1) ASA 150 RF, Flanged as per table B 16.5
2) BSP(M), Threaded (Upto 50NB only)
3) SS Tri-clover (Upto 50NB only)
MountingIn-Line (Horizontal or Vertical)
Operating ConditionsTemp: 0-55°C, Humidity 5-95% non-condensing

Installation Specifications

MediaLiquids (Clear)
Viscosity100 cp max
Line Size15 NB to 150 NB
Display16X2 LCD-4 digit for flow rate & 8 digit for totalized flow (flow rate & totalized flow will be visible through toggle key)
Transmitter ElectronicsOption 1: Integral
Option 2: Remote Remote Electronics Cable: 25mtr
Type of Output1) 4-20mA DC
2) Pulse-30mA Peak-Peak
Calibration RangeAs per requirement (Factory Calibrated)
Accuracy±1% of F.S. (For 20-100% flow)
Linearity±1%
Repeatability±1%
Pressure DropApprox. 0.28 kg/cm² at max flow
Turn down ratio10:1 or 100:1
Process Temperature0-85°C
Max Process Pressure0-10 kg/cm²
Material of constructionBearings: Tungsten Carbide Sleeve/V Jewel
Rotor: SS410/17.PH
Shaft: Tungsten Carbide
Body/Support/Flange: SS
Power Supply24 V DC
Power Consumption<0.5 VA
Response Time<100mSec
Temperature coefficient±0.01% per degree C
Transmitter EnclosureDie Cast Aluminum IP 68
Process Connections1) ASA 150 RF, Flanged as per table B 16.5
2) BSP(M), Threaded (Upto 50NB only)
3) SS Tri-clover (Upto 50NB only)
MountingIn-Line (Horizontal or Vertical)
Operating ConditionsTemp: 0-55°C, Humidity 5-95% non-condensing
DIMENSIONAL DETAILS
Line SizeFlow RangeFlange ANSI150 (B Details 16.5)
m3/hrLPMDCRDBCdFD
15NB0.4 to 4.06.6 to 66.688.911.235,160.515.875mm
20NB0.8 to 8.013.3 to 133.398.612.742.969.915.885mm
25NB1.6 to 16.026.6 to 266.6108.014.250.879.315.8100mm
40NB3.4 to 34.056.6 to 566.6127.017.573.298.615.8140mm
50NB6.8 to 68.0113.0 to 1133.0152.419.190.2120.715.8150mm
100NB13.5 to 135.0225.0 to 2250.0190.523.9127.0152.419.1250mm
150NB27.0 to 270.0450.0 to 4500.0228.623.9157.2190.519.1300mm

INSTALLATION