Electromagnetic Flow Meter

ParameterSpecification
MediaLiquids (Conductive)
Conductivity> 5 µs/cm
Viscosity200 cp max
Line Size15 NB to 1000 NB
ExcitationPulsed DC
Type of Output1) 4 to 20 mA DC, Isolated
2) RS485 MODBUS
3) Pulse
DisplayGraphics type
Engineering UnitUser Programmable (m³/hr by default)
Calibration RangeAs per requirement (Factory Calibrated)
Accuracy+/- 0.5% of F.S. (for 20 to 100% flow)
Linearity+/- 0.5% of F.S.
Repeatability+/- 0.2% of F.S.
Temperature Coefficient+/- 0.05% per °C
Process Temperature85°C max for Rubber Lining & 150°C for PTFE Lining
Process Pressure10 kg/cm² max (higher on request)

 

DESCRIPTION:

Electromagnetic flow meters are mainly used in various industries, ETP, STP, WTP Plants, Borewell and for conductive liquids for measuring the water flow rate. MAG -110 are micro-controller based full bore type electromagnetic flow meters specially used for various industrial applications. These flow meters accurately measures the flow rate of conductive liquid & slurries in closed pipes. Due to simple & rigid design, the flow meter is an obstruction less & maintenance free instrument in place of conventional mechanical flow measuring Device. The use of ‘Pulsed DC technology offers highest ability & better measuring accuracy in the form of electrical signal 4 – 20mA DC linearly proportional to volumetric flow. The instrument is Based on Faraday’s law of electro-magnetic induction. A magnetic field is generated by the instrument in the flow tube. Fluid flowing through this magnetic field generates a voltage that is proportional to the flow velocity. Corresponding electrical output is provided with respect to measuring flow range.

SALIENT FEATURES:

  • Suitable for conductive liquids
  • Maintenance free
  • Full bore type
  • Local Indication through LCD
  • Simple & cost effective construction
  • Empty pipe indication
  • Material option depending upon process data
  • Universal Power Supply 90 to 260V AC Optional 24V DC

TECHNICAL SPECIFICATIONS:

ParameterSpecification
Material of ConstructionLining – Rubber (5mm ±1mm thick) / PTFE (3mm ±1mm thick)
FlangeCS / MS / SS
ElectrodeSS 316L / Hastelloy C / Platinum Coil
HousingMS / SS 304
Power SupplyOption 1: 24 V DC
Option 2: 90 – 260 V AC, 50 Hz
Power Consumption< 10 VA
Isolation1.4 kV between Input, Output & Power Supply
Response Time< 1 Sec
Transmitter EnclosureDie-cast Aluminium IP 66, flow tube IP 68
Process ConnectionsASA 150 flanged, as per table B 16.5
MountingIn-Line Horizontal (Vertical on request)
Operating ConditionsTemperature: 0 to 55°C
Humidity: 5 to 95% non-condensing
Communication PortHART GSM MODEM
ElectronicsRemote Type

MATERIAL OF CONSTRUCTION:

ParameterSpecification
MediaLiquids (Conductive)
Conductivity> 5 µS/cm
Viscosity200 cp max
Line Size15 NB to 1000 NB
ExcitationPulsed DC
Type of Output1. 4 to 20 mA DC, Isolated 2. RS485 MODBUS 3. Pulse
DisplayGraphics type
Engineering UnitUser Programmable (m³/hr by default)
Calibration RangeAs per requirement (Factory Calibrated)
Accuracy±0.5% of F.S. (for 20 to 100% flow)
Linearity±0.5% of F.S.
Repeatability±0.2% of F.S.
Temperature Coefficient±0.05% per °C
Process Temperature85°C max for Rubber Lining
150°C for PTFE Lining
Process Pressure10 kg/cm² max (higher on request)
DIMENSIONS
Meter SizeOD (mm)W (mm)H (mm)FD (mm)MinFlow Range (m3/hr)
NormalMax
15 NB88.91001802000.031.27236.4
20 NB98.41001802000.112.261911
25 NB107.91001802000.183.534118
32 NB117.51002102000.295.790629
40 NB127.01002102000.459.0545
50 NB152.41002102000.7114.1471
65 NB177.81002202001.1923.892119
80 NB190.51002402001.8136.19181
100 NB228.51502742502.8356.55283
125 NB254.01753002504.4288.35442
150 NB279.41753303006.36127.23636
200 NB342.617539035011.3226.181130
250 NB406.424444045017.66353.411766
300 NB482.625052050025.43508.912543
350 NB533.425052055034.62692.683462
400 NB596.225052060045.22904.724522
450 NB635.062363269857.231145.045723
500 NB698.662368676870.651413.637065
600 NB812.6818772918101.742035.6310174

Dimensional Details

Installation 3D drawing